Intraocular lens movement caused by ciliary muscle contraction
Intraocular lens movement caused by ciliary muscle contraction
复制标题
DOI:
10.1016/s0886-3350(02)01652-8
复制
发表时间:
2003-04-01
影响因子:
2.8
通讯作者:
Drexler, W
中科院分区:
文献类型:
--
作者:
Findl, O;Kiss, B;Drexler, W
Purpose: To investigate intraocular lens (IOL) movement, measured as a change in anterior chamber depth (ACD) caused by pilocarpine-induced ciliary muscle contraction.Setting. Department of Ophthalmology, University of Vienna, Vienna, Austria.Methods: In this prospective study, the ACD was measured using high-precision, high-resolution, dual-beam partial coherence interferometry in 62 pseudophakic eyes of 55 patients under pilocarpine- and cyclopentolate-induced ciliary muscle contraction and relaxation. The following were studied: 2 models of a ring-haptic IOL (designed to accommodate), a plate-haptic IOL, and 3 types of 3-piece IOLs. Measurements were performed 3 months after surgery.Results. The ring-haptic IOLs and plate-haptic IOL showed a forward movement (ring haptic 43A, -116 mum; ring haptic 43E, -222 mum; plate haptic -162 mum). The 3-piece IOLs showed no change in ACD except in 1 IOL type in which there was backward movement (156 mum).Conclusions: Pilocarpine-induced ciliary muscle contraction caused forward movement of ring- and plate-haptic IOLs that resulted in an estimated accommodative amplitude of less than 0.50 diopter in most cases. The accommodating ring-haptic IOLs did not perform better than the conventional plate-haptic IOL.