Raster-Mode Continuous-Flow Liquid Microjunction Mass Spectrometry Imaging of Proteins in Thin Tissue Sections
Raster-Mode Continuous-Flow Liquid Microjunction Mass Spectrometry Imaging of Proteins in Thin Tissue Sections
复制标题
DOI:
10.1021/acs.analchem.7b00977
复制
发表时间:
2017-06-06
影响因子:
7.4
通讯作者:
Cooper, Helen J.
中科院分区:
文献类型:
--
作者:
Griffiths, Rian L.;Randall, Elizabeth C.;Cooper, Helen J.
Mass spectrometry imaging by use of continuous-flow liquid micro junction sampling at discrete locations (array mode) has previously been demonstrated. In this Letter, we demonstrate continuous-flow liquid microjunction mass spectrometry imaging of proteins from thin tissue sections in raster mode and discuss advantages (a 10-fold seduction in analysis time) and. challenges (suitable solvent systems, data interpretation) of the approach:: Visualization of data is nontrivial, requiring correlation of solvent-flow, mass spectral data acquisition rate, data quality, and liquid microjunction sampling area. The latter is particularly important for determining optimum pixel size. The minimum achievable pixel size is related to the scan time of the instrument used. Here we show a minimum achievable pixel size of 50 mu m (x-dimension) when using an Orbitrap Elite; however a pixel size of 600 mu m is recommended in order to minimize the effects of oversampling oh image accuracy.