FOURIER-TRANSFORM ION MOBILITY SPECTROMETRY
FOURIER-TRANSFORM ION MOBILITY SPECTROMETRY
复制标题
DOI:
10.1021/ac50001a018
复制
发表时间:
1985-01-01
影响因子:
7.4
通讯作者:
HILL, HH
中科院分区:
文献类型:
--
作者:
KNORR, FJ;EATHERTON, RL;HILL, HH
A mobility Interferogram is generated when the entrance and exit gates of a two-gate ion mobilityspectrometer are si-multaneously opened and closed by a frequency sweeping square wave generator. Fourier transformation of the Inter-ferogram recovers the normal mobility spectrum. A theory for the method is developed, and transformed spectra are obtained for the positive background reactant Ions and for1-methylnaphthalene In N2 drift gas. Potential benefitsof the method for mobility detection In gas chromatography Include rapid collection of complete mobility spectra and increased detector sensitivity. The new method should be generally applicable to other time dispersive techniques such as time of flight mass spectrometry.