A new confocal stigmatic spectrometer for micro-Raman and microfluorescence spectral imaging analysis: Design and applications
A new confocal stigmatic spectrometer for micro-Raman and microfluorescence spectral imaging analysis: Design and applications
复制标题
用于微拉曼和微荧光光谱成像分析的新型共焦象散光谱仪:设计与应用
DOI:
--
复制
发表时间:
1995
期刊:
影响因子:
--
通讯作者:
M. Manfait
中科院分区:
文献类型:
--
作者:
A. Feofanov;S. Sharonov;P. Valisa;E. D. Silva;I. Nabiev;M. Manfait
A new instrument was developed permitting confocal spectral imaging technique to be carried out with a lateral resolution ∼0.3 μm and an axial resolution ∼1.0 μm for specimen areas ranging from 5×5 to 150×150 μm. The modular Raman/fluorescence spectrometer was equipped with a CCD camera, microscope, motorized sample stage, and the confocal entrance chamber. A system of galvanometer controlled mirror scanners equipped the confocal entrance chamber, allowing spectra to be accumulated up to several hundreds of points of sample in parallel, with adjustable spectral and spatial resolution. Stray light rejection property of the Raman spectrometer provides the possibility of the Raman spectral image to be recorded in the low frequency domain. A software was developed to control image accumulation, creation, and treatment. The methods of spectral analysis being applied to a multidimensional set of spectra permit the multiform spectral images to be created. In order to create these images, the different spectral p...