Comparison of the performance characteristics of poly(dimethylsiloxane) and Pyrex microchip electrophoresis devices for peptide separations
Comparison of the performance characteristics of poly(dimethylsiloxane) and Pyrex microchip electrophoresis devices for peptide separations
复制标题
DOI:
10.1016/s0021-9673(03)00722-2
复制
发表时间:
2003-07-04
影响因子:
4.1
通讯作者:
Lunte, SM
中科院分区:
文献类型:
--
作者:
Lacher, NA;de Rooij, NF;Lunte, SM
A comparative study of electrophoretic separations of fluorescently labeled peptides and amino acids on poly(dimethylsiloxane) (PDMS) and Pyrex microchips is presented. The separation parameters for each microchip substrate were compared, including electroosmotic flow, plate numbers, resolution, and limits of detection. The effect of buffer composition on the separation was also investigated. Acceptable separations were obtained for most peptides with both substrates; however, PDMS chips exhibited much lower separation efficiencies and longer analysis times. (C) 2003 Elsevier B.V. All rights reserved.