Prospective respiratory motion correction for coronary MR angiography using a 2D image navigator
Prospective respiratory motion correction for coronary MR angiography using a 2D image navigator
复制标题
DOI:
10.1002/mrm.24280
复制
发表时间:
2013-02-01
影响因子:
3.3
通讯作者:
Botnar, Rene M.
中科院分区:
文献类型:
--
作者:
Henningsson, Markus;Smink, Jouke;Botnar, Rene M.
Respiratory motion remains the major impediment in a substantial amount of patients undergoing coronary magnetic resonance angiography. Motion correction in coronary magnetic resonance angiography is typically performed with a diaphragmatic 1D navigator (1Dnav) assuming a constant linear relationship between diaphragmatic and cardiac respiratory motion. In this work, a novel 2D navigator (2Dnav) is proposed, which prospectively corrects for translational motion in foothead and leftright direction. First, 1Dnav- and 2Dnav-based motion correction are compared in 2D real time imaging experiments, by evaluating the residual respiratory motion in 10 healthy subjects as well as in a moving vessel phantom. Subsequently, 1Dnav and 2Dnav corrected high-resolution 3D coronary MR angiograms were acquired, and both objective and subjective image quality were assessed. For a gating window of 10 mm, 1Dnav and 2Dnav performed equally well; however, without any respiratory gating, the 1Dnav had a lower visual score for all coronary arteries compared with 10 mm gating, whereas the 2Dnav without gating performed similar to 1Dnav with 10 mm gating. Magn Reson Med, 2013. (C) 2012 Wiley Periodicals, Inc.