An electrochemical microsensor for chloride.

An electrochemical microsensor for chloride.
复制标题

氯化物电化学微型传感器。

DOI:
10.1016/0039-9140(91)80017-t
复制
发表时间:
1991
期刊:
影响因子:
6.1
通讯作者:
G. Che
G. Che
中科院分区:
化学1区
文献类型:
--
作者:
S. Dong;G. Che

文献摘要

被引文献

相似文献

研究了碳纤维(CF)上掺杂氯离子的聚吡咯(PPY)膜微盘电极的电位响应。讨论了聚合条件对响应特性的影响。电极的最佳制备条件为:在0.1- 0.2M吡咯(Py)中,含0.1MLiCl,循环电位为+0.8 ~+1.0V,电聚合时间为15-20 min,电极的能斯特响应为56-58 mV/pCl,检出限为3.6 × 10 - 5 MCl。它具有电阻低、调理时间短、响应速度快等优点。用于血清中氯化物的测定,结果满意。
The potential-response of a microdisk electrode made with a chloride-doped polypyrrole (PPY) film on a carbon fibre (CF) has been examined. The effect of the polymerization conditions on the response characteristics is discussed. The optimum conditions for preparing the electrode are: cycling potential from +0.8 to +1.0 V in 0.1–0.2Mpyrrole (Py) containing 0.1MLiCl, electropolymerization time 15–20 min. The electrode gives a Nernstian response of 56–58 mV/pCl and a detection limit of 3.6 × 10−5Mchloride. It has the advantages of low resistance, short conditioning time and fast response. It has been used satisfactorily for detection of chloride in serum.