High Frame Rate Electrical Impedance Tomography System for Monitoring of Regional Lung Ventilation.
High Frame Rate Electrical Impedance Tomography System for Monitoring of Regional Lung Ventilation.
复制标题
用于监测区域肺通气的高帧率电阻抗断层扫描系统。
DOI:
10.1109/embc48229.2022.9871479
复制
发表时间:
2022
期刊:
影响因子:
--
通讯作者:
Rahal M
中科院分区:
文献类型:
--
作者:
Rahal M
This paper describes the development of a compact high frame rate passive electrical impedance tomography system. The injected current amplitude and frequency can be adjusted to fit any EIT application. Measured results show that the system is capable of high frame rate of 89 fps and has power consumption of 1.7 W. It has automatic gain control that reduces noise and improves the quality of the measured EIT image. A comparison is made with other EIT systems to show the potential of the developed system. Clinical Relevance— The developed EIT system has application in the clinical assessment of neonatal and SARS-Co V-2 patients. In these applications there is an urgent need for a low cost bedside non-invasive imaging system to continuously monitor dynamic changes in regional lung ventilation