An improved flow microfluorometer for rapid measurement of cell fluorescence.
An improved flow microfluorometer for rapid measurement of cell fluorescence.
复制标题
一种改进的流动微荧光计,用于快速测量细胞荧光。
DOI:
10.1016/0014-4827(73)90280-2
复制
发表时间:
1973
影响因子:
3.7
通讯作者:
L. Cram
中科院分区:
文献类型:
--
作者:
D. Holm;L. Cram
An improved instrument for the making of high-speed fluorescent measurements on single cells has been constructed and characterized. Instrumental optimization was accomplished with Chinese hamser cells (line CHO) stained by the fluorescent Feulgen procedure using either auramine-0 or acriflavine. Contributions of instrumental resolution and artifacts to cellular DNA distributions have been determined. The instrument introduces a total coefficient of variation of less than 2% for the case of CHO cells stained with bright dyes such as acriflavine, where photon-statistical effects are minimal. The design of the instrument is such that any of a number of cellular constituents or properties can be studied and quantitated on a single-cell basis at a rate of 50 000 cells/min.