Breast implant rupture is an important complication of augmented and reconstructed breasts. Although several techniques such as mammography, xeromammography, ultrasound, thermography, and computed tomographic (CT) scanning have been proven to be useful to detect implant rupture, they have several disadvantages and lack specificity. In the current study, we have established magnetic resonance imaging (MRI) as a definitive, reliable, and reproducible technique to diagnose both intracapsular and extracapsular ruptures. The study was conducted in 100 symptomatic patients. Our imaging parameters were able to identify ruptures in implants with silicone shells. All the ruptures showed the presence of wavy lines, free-floating silicone shell within the gel (“free-floating loose-thread sign” or “linguine sign”). We had a 3.75 percent incidence of false-positive and false-negative results. The sensitivity for detection of silicone implant rupture was 76 percent, with a specificity of 97 percent. In addition, we also were able to identify the artifacts that may interfere with the definitive diagnosis of implant rupture. (Plast. Reconstr. Surg. 92: 681, 1993.)