Limitations of single-channel EEG on the forehead for neonatal seizure detection
Limitations of single-channel EEG on the forehead for neonatal seizure detection
复制标题
DOI:
10.1038/jp.2008.195
复制
发表时间:
2009-03-01
影响因子:
2.9
通讯作者:
Clancy, R. R.
中科院分区:
文献类型:
--
作者:
Wusthoff, C. J.;Shellhaas, R. A.;Clancy, R. R.
Objectives: In amplitude-integrated EEG, lead placement across the forehead is convenient, but this location has unknown effects on neonatal seizure (NS) detection sensitivity. This study describes the limits of NS detection by a single forehead EEG channel.Study Design: Records were taken from a digital library of conventional EEGs (CEEGs) with NS, previously characterized at a bicentral channel, C(3) -> C(4). We analyzed electrographic characteristics in a single forehead channel, Fp(3) -> Fp(4).Result: A total of 330 seizures from 125 CEEGs were included. With Fp(3) -> Fp(4) at least one NS was detected in 66% of records vs 90% using C(3) -> C(4) (P < 0.0001). Of 330 seizures, 46% appeared in Fp(3) -> Fp(4) vs 73% in C(3) -> C(4) (P < 0.0001). Seizures appeared briefer in Fp(3) -> Fp(4) than C(3) -> C(4) (P < 0.006) and CEEG (P < 0.0001).Conclusion: NSs are significantly more difficult to detect with a single forehead channel than bicentrally or on CEEG. In Fp(3) -> Fp(4), a third of records with seizures were missed and over half of seizures were undetected.