Computational fluid dynamics analysis of a Maglev centrifugal left ventricular assist device
Computational fluid dynamics analysis of a Maglev centrifugal left ventricular assist device
复制标题
DOI:
10.1111/j.1525-1594.2004.07384.x
复制
发表时间:
2004-10-01
影响因子:
2.4
通讯作者:
Schöb, R
中科院分区:
文献类型:
--
作者:
Burgreen, GW;Loree, HM;Schöb, R
The fluid dynamics of the Thoratec HeartMate III (Thoratec Corp., Pleasanton, CA, U.S.A.) left ventricular assist device are analyzed over a range of physiological operating conditions. The HeartMate III is a centrifugal flow pump with a magnetically suspended rotor. The complete pump was analyzed using computational fluid dynamics (CFD) analysis and experimental particle imaging flow visualization (PIFV). A comparison of CFD predictions to experimental imaging shows good agreement. Both CFD and experimental PIFV confirmed well-behaved flow fields in the main components of the HeartMate III pump: inlet, volute, and outlet. The HeartMate III is shown to exhibit clean flow features and good surface washing across its entire operating range.