[Investigation for bone fixation effect of thin HA coated layer on Ti implants].
[Investigation for bone fixation effect of thin HA coated layer on Ti implants].
复制标题
HA薄层对钛种植体骨固定效果的研究
DOI:
10.5357/koubyou.72.247
复制
发表时间:
2005
期刊:
影响因子:
--
通讯作者:
A. Fujisawa
中科院分区:
文献类型:
--
作者:
A. Fujisawa
In 1995, the author reported the development of unique cementless total hip joints applying hydroxyapatite (HA) flame coating onto a porous titanium surface which is arc-sprayed with pure titanium material in argon gas at atmospheric pressure. The surface roughness of the porous layer was 360 microm Rmax, and the layer was more resistant to blast erosion than the conventional low pressure plasma sprayed porous layers. The thickness of the HA layer was between 20 and 40 microns considering the balance of bone conduction effect of HA and the risk of mechanical detachment of the layers from the porous titanium. Short-term animal tests showed excellent results. At that time a specific question was raised as to whether or not the coated HA layers would remain in the living bone for a long time. The results obtained from observation of retrieved specimens show that the thickness of the HA coating layer is sufficient for new bone formation after implantation, and HA would be absorbed in the body within a few years, after contributing to the new bone formation.
DOI:
10.2106/00004623-198668090-00013
发表时间:
1986-12
期刊:
The Journal of bone and joint surgery. American volume
影响因子:
--
作者:
T. Turner;D. Sumner;R. Urban;D. Rivero;J. Galante
通讯作者:
T. Turner;D. Sumner;R. Urban;D. Rivero;J. Galante