MR coronary angiography in pigs with intraarterial injections of a hyperpolarized 13C substance
MR coronary angiography in pigs with intraarterial injections of a hyperpolarized 13C substance
复制标题
动脉内注射超极化 13C 物质对猪进行 MR 冠状动脉造影
DOI:
10.1002/mrm.20847
复制
发表时间:
2006
影响因子:
3.3
通讯作者:
J. Petersson
中科院分区:
文献类型:
--
作者:
L. Olsson;C. Chai;O. Axelsson;M. Karlsson;K. Golman;J. Petersson
A new diagnostic application of a water‐soluble contrast medium (CM) based on the hyperpolarization of a 13C substance is introduced. The degree of polarization achieved is >30%, which is about a factor of 105 higher than the thermal equilibrium polarization level at 1.5 T. Imaging of hyperpolarized (HP) CM during a cardiac interventional MRI procedure was studied. Catheters were positioned in the left and right coronary arteries of pigs. A coil tuned to 13C was used for nonproton imaging. The HP‐13C CM (∼5 ml, 0.5 M, ∼30% polarization) was injected during projection imaging using a fully balanced steady‐state free precession (SSFP) pulse sequence with and without cardiac gating. The contrast agent‐filled catheter was clearly visible during the procedure. The coronary arteries were well depicted and the signal‐to‐noise ratios (SNRs) were in the range of 10–40. The use of HP‐13C CM may provide a new diagnostic procedure for interventional MRI. Magn Reson Med, 2006. © 2006 Wiley‐Liss, Inc.