Determination of 8-epi PGF2α concentrations as a biomarker of oxidative stress using triple-stage liquid chromatography/tandem mass spectrometry

Determination of 8-epi PGF2α concentrations as a biomarker of oxidative stress using triple-stage liquid chromatography/tandem mass spectrometry
复制标题

DOI:
10.1002/rcm.4197
复制
发表时间:
2009-09-01
影响因子:
2
通讯作者:
Blomhoff, Rune
Blomhoff, Rune
中科院分区:
化学3区
文献类型:
--
作者:
Bastani, Nasser E.;Gundersen, Thomas E.;Blomhoff, Rune

文献摘要

被引文献

相似文献

F2-异前列腺素是一类前列腺素F2样化合物,由花生四烯酸的自由基催化过氧化反应形成。几种F2-异前列烷,但特别是8-表PGF(2 α),被广泛用作氧化应激生物标志物。建立了一种基于液相色谱-电喷雾负离子(ESI)串联质谱检测(LC/MS/MS)的分析方法,用于测定人血浆、全血、红细胞和尿液中的8-epi PGF(2 α)浓度。8-表PGF(2 α)-d4,8-表PGF(2 α)的稳定同位素衍生物,用作内标(IS)。将50 μ L样品聚焦在柱上,并在两个串联连接的3 μ m粒径SUPELCOSIL(TM)ABZ+Plus HPLC柱(15 cm × 4.6 mm和7.5 cm × 4.6 mm)上分离。具有ESI的Applied Biosystems 4000 Q TRAP LC/MS/MS系统以多反应监测(MRM)模式操作,其中吸附剂至产物离子跃迁m/z 353.4 -> 193.1(8-差向PGF(2 α))、357.4 -> 197.1(8-差向PGF(2 α)-d4)用于定量。对该试验进行了充分验证,发现其具有足够的准确度、精密度、线性、灵敏度和选择性。检测质量限(mLOD)为1 pg从色谱柱洗脱的分析物。该方法已成功地应用于人血浆、全血、红细胞和尿液样品的分析。版权所有(C)2009约翰威利父子有限公司
F2-isoprostanes are a family of prostaglandin F2-like compounds that are formed by free-radical-catalyzed peroxidation of arachidonic acid. Several F2-isoprostanes, but in particular 8-epi PGF(2 alpha), are widely used as oxidative stress biomarkers. An analytical method based on liquid chromatography with negative electrospray ionization (ESI) coupled to tandem mass spectrometric detection (LC/MS/MS) was developed for the determination of 8-epi PGF(2 alpha) concentrations in human plasma, whole blood, erythrocytes and urine. 8-epi PGF(2 alpha)-d4, a stable isotope derivative of 8-epi PGF(2 alpha), was used as an internal standard (IS). A 50 mu L sample was focused on-column and separated on two 3 mu m particle size SUPELCOSIL (TM) ABZ+Plus HPLC columns (15 cm x 4.6 mm and 7.5 cm x 4.6 mm) connected in series. An Applied Biosystems 4000 Q TRAP LC/MS/MS system with ESI was operated in multiple reaction monitoring (MRM) mode with the precursor-to-product ion transitions m/z 353.4 -> 193.1 (8-epi PGF(2 alpha)), 357.4 -> 197.1 (8-epi PGF(2 alpha)-d4), used for quantification. The assay was fully validated and found to have adequate accuracy, precision, linearity, sensitivity and selectivity. The mass limit of detection (mLOD) was 1 pg of analyte eluting from the column. The assay has been successfully applied to the analysis of human plasma, whole blood, erythrocytes and urine samples. Copyright (C) 2009 John Wiley & Sons, Ltd.