Some characteristics of a protein-coated ODS column and its use for the determination of drugs by the direct injection analysis of plasma samples
Some characteristics of a protein-coated ODS column and its use for the determination of drugs by the direct injection analysis of plasma samples
复制标题
蛋白质包被 ODS 柱的一些特性及其在血浆样品直接进样分析药物测定中的应用
DOI:
10.1007/bf02687789
复制
发表时间:
1984
期刊:
影响因子:
1.7
通讯作者:
H. Imai
中科院分区:
文献类型:
--
作者:
Hisanobu Yoshida;I. Morita;G. Tamai;T. Masujima;T. Tsuru;N. Takai;H. Imai
It was found that an ODS column of small pore (120Å) which was coated with denatured plasma proteins (protein-coated ODS) no longer adsorbed plasma proteins from aqueous solution but retained the characteristics of native ODS for small hydrophobic molecules. Elemental analysis and nitrogen desorption (BET) analysis showed that the protein-coated ODS contained ca 25 mg proteing−1dry resin and that the pore diameter or pore volume was similar to that of native ODS. The coated denatured proteins, which seemed to be adsorbed on the external surface of the porous resin, were not eluted under usual reverse-phase elution conditions. Operating as either an analytical column or a pre-column, this protein-coated ODS column was used to analyse spiked-drugs in plasma. The recovery of all the spiked-drugs (such as doxorubicin, methotrexate) was almost quantitative (98–102%) with good reproducibilities (c.v., less than 2%). The present method was useful for the determination of total, that is, free + bound-to-plasma-proteins, hydrophobic drugs in plasma in view of its accuracy and simplicity.