A NEW STEADY-STATE IMAGING SEQUENCE FOR SIMULTANEOUS ACQUISITION OF 2-MR IMAGES WITH CLEARLY DIFFERENT CONTRASTS

A NEW STEADY-STATE IMAGING SEQUENCE FOR SIMULTANEOUS ACQUISITION OF 2-MR IMAGES WITH CLEARLY DIFFERENT CONTRASTS
复制标题

DOI:
10.1002/mrm.1910070105
复制
发表时间:
1988-05-01
影响因子:
3.3
通讯作者:
DEIMLING, M
DEIMLING, M
中科院分区:
医学3区
文献类型:
--
作者:
BRUDER, H;FISCHER, H;DEIMLING, M

文献摘要

被引文献

相似文献

我们提出了一种新的稳态成像序列,它同时允许在一个单一的收购形成两个MR图像具有明显不同的对比度。第一图像的对比度是FISP样的,而第二图像是强T2加权的。原则上,图像中的T2值可以从第一图像和第二图像的组合来计算。我们还显示了计算的T2图像。
We present a new steady-state imaging sequence, which simultaneously allows in a single acquisition the formation of two MR images with clearly different contrasts. The contrast of the first image is FISP-like, whereas the second image is strongly T2-weighted. In principle the T2 values in the image can be calculated from the combination of the first and second images. We also show calculated T2 images.