Continuous ejection fraction estimation by model-based analysis of an aortic pressure waveform: comparison to echocardiography.

Continuous ejection fraction estimation by model-based analysis of an aortic pressure waveform: comparison to echocardiography.
复制标题

通过基于模型的主动脉压力波形分析来估计连续射血分数:与超声心动图的比较。

DOI:
10.1109/iembs.2007.4352452
复制
发表时间:
2007
期刊:
Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference
影响因子:
--
通讯作者:
Mukkamala,Ramakrishna
Mukkamala,Ramakrishna
中科院分区:
--
文献类型:
--
作者:
Swamy,Gokul;Olivier,Bari;Kuiper,Jacob;Mukkamala,Ramakrishna

文献摘要

相似文献

Left ventricular ejection fraction (EF) is perhaps the most clinically significant index of global ventricular function. EF is measured in clinical practice using imaging methods such as non-invasive echocardiography. However, imaging methods generally require a skilled operator and expensive equipment. Thus, EF is not sufficiently monitored. To this end, we have recently developed a novel technique to continuously (i.e., automatically) estimate EF by model-based analysis of an aortic pressure waveform. Here, we review the technique and present its evaluation with respect to reference echocardiography measurements from three dogs during diverse interventions. We report an overall EF error of only 8.3%. With further successful testing, the technique may ultimately be utilized for continuous EF monitoring in research and clinical settings in which an aortic catheter is employed.