Risk Analysis Methods for Complying with Z540.3
Risk Analysis Methods for Complying with Z540.3
复制标题
遵守Z540.3的风险分析方法
DOI:
--
复制
发表时间:
2008
期刊:
影响因子:
--
通讯作者:
J. Mayer
中科院分区:
文献类型:
--
作者:
S. Q. Wang;J. Mayer
Abstract Clause 5.3 of ANSI/NCSL Z540.3, "Requirements for the Calibration of Measuring and Test Equipment," contains a risk management requirement for cases where the results of calibration are employed to verify that quantities are within specified tolerances. This requirement limits the probability of incorrect acceptance decisions (false accept risk) to a maximum of 2%. Where the estimation of this probability is not feasible, the test uncertainty ratio (TUR) is required to be 4:1 or greater.