Accuracy of Quantification of Iodine and Hounsfield Unit Values on Virtual Monochromatic Imaging Using Dual-Energy Computed Tomography
Accuracy of Quantification of Iodine and Hounsfield Unit Values on Virtual Monochromatic Imaging Using Dual-Energy Computed Tomography
复制标题
双能计算机断层扫描虚拟单色成像中碘和亨斯菲尔德单位值定量的准确性
DOI:
10.1097/rct.0000000000000798
复制
发表时间:
2018
影响因子:
1.3
通讯作者:
Teshima Teruki
中科院分区:
文献类型:
--
作者:
Washio Hayate;Ohira Shingo;Karino Tsukasa;Nitta Yuya;Hayashi Masaaki;Miyazaki Masayoshi;Teshima Teruki
ObjectiveThe aim of the study was to compare the accuracy of quantification of iodine and Hounsfield unit (HU) values on virtual monochromatic imaging (VMI) using dual-layer computed tomography (DLCT) and fast kilovolt-switching computed tomography (FKSCT).Materials and MethodsThis study was performed in 2 phantoms (large and small) using 16 rods representing different materials (iodine, calcium, blood, and adipose tissue) with different dimensions and concentrations. The absolute percentage errors (absolute ratio of measurement error to true iodine concentration) for iodine concentration and HU value on VMI at 50, 70, and 100 keV were compared between DLCT and FKSCT. The Mann–Whitney U test was used to assess statistical significance.ResultsOverall, the absolute percentage errors for iodine concentration and HU value on VMI were smaller for DLCT than for FKSCT.ConclusionsOverall, the accuracy of iodine and HU values was higher for DLCT than for FKSCT.