Continuous intraoperative neuromonitoring in thyroid surgery: Safety analysis of 400 consecutive electrode probe placements with standardized procedures
Continuous intraoperative neuromonitoring in thyroid surgery: Safety analysis of 400 consecutive electrode probe placements with standardized procedures
复制标题
DOI:
10.1002/hed.24280
复制
发表时间:
2016-04-01
影响因子:
2.9
通讯作者:
Dionigi, Gianlorenzo
中科院分区:
文献类型:
--
作者:
Mangano, Alberto;Kim, Hoon Yub;Dionigi, Gianlorenzo
Background. Continuous intraoperative neuromonitoring (C-IONM) is a new technology and it is appropriate to analyze its safety.Methods. C-IONM was performed according to a standardized technique to control any adverse events and electrode positioning issues.Results. Four hundred vagal nerve dissections were analyzed considering vagal nerve diameter, mean time effort for C-IONM probe positioning, and electrode dislocation rate. A significant superior dislocation rate in case of: (a) when a 3 mm automatic periodic stimulating (APS) electrode size was used in a vagal nerve diameter