Recording compliance of dental splint use in obstructive sleep apnoea patients by force and temperature modelling
Recording compliance of dental splint use in obstructive sleep apnoea patients by force and temperature modelling
复制标题
DOI:
10.1007/bf02344801
复制
发表时间:
2001-03-01
影响因子:
3.2
通讯作者:
Hao, J
中科院分区:
文献类型:
--
作者:
Tjin, SC;Tan, YK;Hao, J
Fibre-optic sensors are used to monitor the force and temperature of dental splints worn by patients suffering from sleep apnoea. Owing to the small sire of the sensors, they can be easily embedded within the splint in a way that does not affect the effectiveness of the splint, and, at the same time, are able to indicate whether the splint has been properly worn by the patient. The overall dimensions of the sensor are approximately 0.375 mm thickness, 1 cm length and 3 mm width. The force and temperature sensors are calibrated and found to have sensitivities of better than 0.5 N and 0.1 degreesC, respectively. Trials performed on patients show that the measurement of pressure and temperature is an effective way of monitoring the proper usage of the dental splint by the patients.