Evaluation of a matched filter resolution recovery reconstruction algorithm for SPECT-CT imaging
Evaluation of a matched filter resolution recovery reconstruction algorithm for SPECT-CT imaging
复制标题
DOI:
10.1097/mnm.0b013e32835ce5b5
复制
发表时间:
2013-03-01
影响因子:
1.5
通讯作者:
Murray, Iain
中科院分区:
文献类型:
--
作者:
O'Mahoney, Eoin;Murray, Iain
Aim The aim of our study was to assess improvements in spatial resolution and noise control from the application of the Astonish resolution recovery algorithm for single photon emission computed tomography imaging. Secondary aims were to compare acquisitions made with low-energy general purpose collimators with those obtained using low-energy high-resolution collimators in this context and evaluate the potential of a finer matrix to improve image quality further.Materials and methods A Tc-99m-filled Jaszczak phantom with hot spheres was used to assess contrast and noise. A National Electrical Manufacturers Association triple line source single photon emission computed tomography resolution phantom was used to measure spatial resolution. Acquisitions were made using both low-energy high-resolution and low-energy general purpose collimators.Results Compared with standard ordered subsets expectation maximization reconstructions, the resolution recovery algorithm resulted in a higher spatial resolution (8 vs. 14 mm full-width at half-maximum) leading to reduced partial volume effects in the smaller Jaszczak spheres. Higher image contrast was achieved alongside lower levels of noise. An edge enhancement artefact was observed in the resolution recovery corrected images. An overestimate of the target-to-background activity was also observed for the larger spheres.Conclusion The use of such an algorithm results in images characterized by increased spatial resolution and reduced noise. However, small sources of the order of 2-3cm can be significantly overenhanced. Nucl Med Commun 34: 240-248 (C) 2013 Wolters Kluwer Health vertical bar Lippincott Williams & Wilkins.