Evaluation of a new system for self blood glucose monitoring.

Evaluation of a new system for self blood glucose monitoring.
复制标题

自我血糖监测新系统的评估。

DOI:
10.1016/s0168-8227(88)80020-2
复制
发表时间:
1988
影响因子:
5.1
通讯作者:
Simineiro,LM
Simineiro,LM
中科院分区:
医学3区
文献类型:
--
作者:
Clarke,WL;Becker,DJ;Cox,D;Santiago,JV;White,NH;Betschart,J;Eckenrode,K;Levandoski,LA;Prusinski,EA;Simineiro,LM

文献摘要

被引文献

相似文献

设计了一种新型组合反射仪/目视判读系统(Glucometer II®/Glucostix®,艾姆斯分部,Miles Laboratories,埃尔克哈特,印第安纳州),用于自我血糖监测。对该系统的性能评价表明,血糖仪测得的血糖值与实验室测得的全血血糖值之间存在线性关系(y = 0.95x +2.86,r = 0.97)。此外,95%的目视判读血糖值在YSI比较值的一个色块内。误差网格分析是一种用于确定患者测定血糖结果的临床准确性的新方法,表明该新系统的组件可产生临床准确的血糖结果。
A new combination reflectance meter/visually interpretable system (Glucometer II®/Glucostix®, Ames Division, Miles Laboratories, Elkhart, IN) has been designed for self blood glucose monitoring. Performance evaluation of this system demonstrates a linear relationship between meter-determined blood glucose values and laboratory-determined whole blood glucose values (y = 0.95x + 2.86,r= 0.97). In addition, 95% of visually interpreted blood glucose values are within one color block of YSI comparative values. Error grid analysis, a new method for determining the clinical accuracy of patient-determined blood glucose results, demonstrated that components of this new system produce clinically accurate blood glucose results.