Evaluation of a new system for self blood glucose monitoring.
Evaluation of a new system for self blood glucose monitoring.
复制标题
自我血糖监测新系统的评估。
DOI:
10.1016/s0168-8227(88)80020-2
复制
发表时间:
1988
影响因子:
5.1
通讯作者:
Simineiro,LM
中科院分区:
文献类型:
--
作者:
Clarke,WL;Becker,DJ;Cox,D;Santiago,JV;White,NH;Betschart,J;Eckenrode,K;Levandoski,LA;Prusinski,EA;Simineiro,LM
A new combination reflectance meter/visually interpretable system (Glucometer II®/Glucostix®, Ames Division, Miles Laboratories, Elkhart, IN) has been designed for self blood glucose monitoring. Performance evaluation of this system demonstrates a linear relationship between meter-determined blood glucose values and laboratory-determined whole blood glucose values (y = 0.95x + 2.86,r= 0.97). In addition, 95% of visually interpreted blood glucose values are within one color block of YSI comparative values. Error grid analysis, a new method for determining the clinical accuracy of patient-determined blood glucose results, demonstrated that components of this new system produce clinically accurate blood glucose results.