Validation of aerosport KB1-C portable metabolic system
Validation of aerosport KB1-C portable metabolic system
复制标题
DOI:
10.1055/s-2007-971135
复制
发表时间:
1999-07-01
影响因子:
2.5
通讯作者:
Ainsworth, BE
中科院分区:
文献类型:
--
作者:
King, GA;McLaughlin, JE;Ainsworth, BE
The purpose of this study was to test the validity of the Aerosport KB1-C portable metabolic measurement system against the criterion Douglas bag method. During cycle ergometry, simultaneous measurements of minute ventilation ((V) over dot (E)), oxygen consumption ((V) over dot (2)), and carbon dioxide production ((V) over dot CO2) were made at rest and at power outputs of 50, 100, 150, 200, and 250 W. No significant differences (P > 0.05) were observed for (V)over dot (E), (V) over dot O (2), and (V) over dot CO2 between the KB1-C and Douglas bag at 100, 150, and 250 W, while the KB1-C was significantly different (P