Accuracy assessment of a novel blood pressure measurement device in a South African adult population: Tensoval duo control

Accuracy assessment of a novel blood pressure measurement device in a South African adult population: Tensoval duo control
复制标题

DOI:
10.1097/mbp.0b013e32834db648
复制
发表时间:
2011-12-01
影响因子:
1.3
通讯作者:
Shennan, Andrew H.
Shennan, Andrew H.
中科院分区:
医学4区
文献类型:
--
作者:
de Greeff, Annemarie;Hezelgrave, Natasha;Shennan, Andrew H.

文献摘要

被引文献

相似文献

目的Tensoval DUO控制仪采用听诊和示波测量技术相结合的新技术来测量血压。我们根据欧洲高血压学会的国际协议在南非成年人群中评估了该装置的准确性。方法获得伦理批准,从金伯利医院综合体(南非)招募33名参与者。结果Tensoval Duo控制仪通过了方案的所有阶段,在收缩压和舒张压方面的平均标准偏差分别为-3.2+/-7.2和2.0+/-6.0 mm Hg。结论本研究证实了Tensoval Duo控制仪在成人中的准确性和临床应用建议。在特殊患者群体中的准确性还有待评估,但这项新技术可能对那些患有血管疾病的人特别有价值。血液出版社Monit 16:304-306(C)2011 Wolters Kluwer Health|Lippincott Williams&Wilkins。血压监测2011,16:304-306
Objective The Tensoval duo control device uses a novel combination of auscultatory and oscillometric technology to measure blood pressure. We assessed the accuracy of this device in a South African adult population according to the International Protocol of the European Society of Hypertension.Methods Ethical approval was obtained and 33 participants were recruited from Kimberley Hospital Complex (South Africa). Trained observers took sequential same arm blood pressure measurements from each participant, alternating between mercury sphygmomanometery and the device.Results The Tensoval duo control passed all the phases of the protocol and achieved a mean difference +/- SD of -3.2 +/- 7.2 and 2.0 +/- 6.0 mmHg for systolic and diastolic pressures, respectively.Conclusion This study confirms the accuracy and recommendation for clinical use of the Tensoval duo control device in adults. Accuracy in special patient groups is yet to be assessed, but this novel technology could be particularly valuable in those with vascular disease. Blood Press Monit 16: 304-306 (C) 2011 Wolters Kluwer Health | Lippincott Williams & Wilkins. Blood Pressure Monitoring 2011, 16: 304-306