A goniometric glove for clinical hand assessment - Construction, calibration and validation

A goniometric glove for clinical hand assessment - Construction, calibration and validation
复制标题

DOI:
10.1054/jhsb.1999.0360
复制
发表时间:
2000-04-01
期刊:
JOURNAL OF HAND SURGERY-BRITISH AND EUROPEAN VOLUME
影响因子:
--
通讯作者:
Harley, P
Harley, P
中科院分区:
其他
文献类型:
--
作者:
Williams, NW;Penrose, JMT;Harley, P

文献摘要

被引文献

相似文献

描述了测角手套的结构,手套中的每个传感器都可以使用已知角度的块作为角度参考在定制的金属手上进行校准。手套每个传感器输出的数字数据被转换为每个关节处的角位移。该手套经过定制金属夹具和人手测量的一致性和准确性进行了验证。发现手套的准确性在传统测角法的范围内。建议测角手套可用于评估手部功能。
The construction of a goniometric glove is described, Each of the sensors in the glove mas calibrated over a custom built metal hand using blocks of known angles as angular references. The digital data output from each sensor of the glove were converted into angular displacements at each joint. The glove was validated for consistency of measurement and accuracy over a custom built metal jig and in the human hand. The accuracy of the glove was found to be within the limits of traditional goniometry. It is proposed that goniometric gloves could be useful in the assessment of hand function.