Contrast in the transmission mode of a low‐voltage scanning electron microscope
Contrast in the transmission mode of a low‐voltage scanning electron microscope
复制标题
低压扫描电子显微镜透射模式对比
DOI:
10.1111/j.1365-2818.1994.tb03444.x
复制
发表时间:
1994
影响因子:
2
通讯作者:
Ludwig Reimer
中科院分区:
文献类型:
--
作者:
U. Golla;B. Schindler;Ludwig Reimer
The contrast thicknesses (xk) of thin carbon and platinum films have been measured in the transmission mode of a low‐voltage scanning electron microscope for apertures of 40 and 100 mrad and electron energies (E) between 1 and 30 keV. The measured values overlap with those previously measured for E (≥ 17keV) in a transmission electron microscope. Differences in the decrease of xk with decreasing E between carbon and platinum agree with Wentzel‐Kramer‐Brillouin calculations of the elastic cross‐sections. Knowing the value of xk allows the exponential decrease ∝ exp(—x/xk) in transmission with increasing mass‐thickness (x = ρt) of the specimen and the increasing gain of contrast for stained biological sections with decreasing electron energy to be calculated for brightfield and darkfield modes.