Technical Note: Estimation of the Damping Coefficient of Fluid‐Filled, Catheter‐Transducer Pressure‐Measuring Systems
Technical Note: Estimation of the Damping Coefficient of Fluid‐Filled, Catheter‐Transducer Pressure‐Measuring Systems
复制标题
技术说明:液体填充导管传感器压力测量系统阻尼系数的估计
DOI:
--
复制
发表时间:
1988
期刊:
影响因子:
--
通讯作者:
J. Bourland
中科院分区:
文献类型:
--
作者:
L. Geddes;J. Bourland
A simple technique is presented that allows determination of the degree of damping and the damped resonant frequency of catheter-transducer systems. From these data, the hydraulic sinusoidal frequency response, as well as the undamped (natural) frequency can be determined. The technique consists of suddenly releasing pressure in a bubble-free, fluid-filled catheter-transducer system while recording pressure with a stripchart recorder.