Basic principles for measurement of intramuscular pressure.

Basic principles for measurement of intramuscular pressure.
复制标题

DOI:
10.1016/s1060-1872(95)80022-0
复制
发表时间:
1995-10-01
影响因子:
0.3
通讯作者:
Ballard, R E
Ballard, R E
中科院分区:
医学4区
文献类型:
--
作者:
Hargens, A R;Ballard, R E

文献摘要

被引文献

相似文献

我们回顾了测量人类肌内压力(IMP)的历史和方法学方法。这些技术提供了有价值的测量肌肉张力和活动的方法,以及评估肌间室综合征的诊断标准。虽然灯芯和导管技术提供了静止状态下IMP的精确测量,但由于低频响应和流体静力(静态和惯性)压力伪影,它们在运动研究和运动室综合征诊断中的价值有限。目前,在动态运动中诊断运动间室综合征的大多数信息是通过my压迫导管获得的。然而,未来的研究和临床诊断可以通过使用微型传感器尖端导管如Millar micro -tip来优化IMP。
We review historical and methodological approaches to measurements of intramuscular pressure (IMP) in humans. These techniques provide valuable measures of muscle tone and activity as well as diagnostic criteria for evaluation of exertional compartment syndrome. Although the wick and catheter techniques provide accurate measurements of IMP at rest, their value for exercise studies and diagnosis of exertional compartment syndrome is limited because of low frequency response and hydrostatic (static and inertial) pressure artifacts. Presently, most information on diagnosis of exertional compartment syndromes during dynamic exercise is available using the Myopress catheter. However, future research and clinical diagnosis using IMP can be optimized by the use of a miniature transducer-tipped catheter such as the Millar Mikro-tip.