A comparison of manual and controlled-force attachment-level measurements.

A comparison of manual and controlled-force attachment-level measurements.
复制标题

手动和受控力附件水平测量的比较。

DOI:
10.1111/j.1600-051x.1997.tb01212.x
复制
发表时间:
1997
影响因子:
6.7
通讯作者:
Geurs,NC
Geurs,NC
中科院分区:
医学1区
文献类型:
--
作者:
Reddy,MS;Palcanis,KG;Geurs,NC

文献摘要

被引文献

相似文献

This study compared the intra‐examiner and inter‐examiner error of 2 constant force probes to the reading of a conventional manual probe, 3 examiners made repeated examinations of attachment level using a modified Florida probe and a manual North Carolina probe (read to 1 mm or 0.5 mm); relative attachment level measurements were made using a Florida disk probe. One probe was used in each quadrant in 8 subjects with moderate to advanced periodontitis. Error was calculated as the mean of the absolute value of the difference between each examination, and the correlation between values at each examination calculated. Statistically‐significant differences between probe type, examiners, and sites were detected using a repeated measures ANOVA accounting for the nesting within subjects. There was a significant difference in error by probe type (modified Florida probe 0.62±0.03 mm, r=0.86; Florida stent probe 0.55±0.05 mm, r= 0.82; manual probe to 1 mm 0.39±0.02 mm, r=0.88; manual probe to 0.5 mm 0.40±0.02 mm, r=0.89;(p<0.001). Significant differences were observed by examiners (p<0.01). These data indicate that both manual and controlled‐force probes can provide measurement within less than 1 mm of error; however, individual calibration of examiners remains important in the reduction of error.