The second generation HRRT - a multi-centre scanner performance investigation
The second generation HRRT - a multi-centre scanner performance investigation
复制标题
第二代 HRRT - 多中心扫描仪性能调查
DOI:
--
复制
发表时间:
2005
期刊:
影响因子:
--
通讯作者:
D. Wong
中科院分区:
文献类型:
--
作者:
V. Sossi;H. de Jong;W. Barker;P. Bloomfield;Z. Burbar;M. Camborde;C. Comtat;L. Eriksson;S. Houle;D. Keator;C. Knob;R. Krais;A. Lammertsma;A. Rahmim;M. Sibomana;M. Teras;C. Thompson;R. Trébossen;J. Votaw;M. Walker;K. Wienhard;D. Wong
The high resolution research tomograph (HRRT) is one of the most complex existing positron emission tomographs: it is the only human size scanner capable of decoding the depth of the /spl gamma/-ray interaction in the crystal, using a lutetium LSO/LYSO phoswich detector arrangement. In this study we determined basic scanner hardware characteristics, such as scanner data acquisition stability, and their variability across eleven centres. In addition a subset of the NEMA NU-2001 standards measurements was performed. We found (i) significant variability in the DOI decoding results between centres, (ii) a trend toward an increasing number of detected true coincident events as a function of elapsed time from scanner calibration likely due to a shifting energy spectrum, (iii) a count-rate dependent layer identification, (iv) scatter fraction ranging from /spl sim/ 42% to 54% where the variability was partly related to the shifting of the energy spectrum, (v) sensitivity ranging from /spl sim/5.5% to 6.5% across centres, (vi) resolution of /spl sim/(2.5 mm)/sup 3/, fairly consistent across centres, (vii) image quality which is very comparable to other scanners.