Improved measurements of thoron (220Rn) in natural waters
Improved measurements of thoron (220Rn) in natural waters
复制标题
DOI:
10.1007/s10967-018-6110-z
复制
发表时间:
2018-08
影响因子:
1.6
通讯作者:
S. Chanyotha;P. Sola;R. Kritsananuwat;D. Lane-Smith;W. Burnett
中科院分区:
文献类型:
--
作者:
S. Chanyotha;P. Sola;R. Kritsananuwat;D. Lane-Smith;W. Burnett
Thoron (220Rn) in water measurements can be useful for prospecting for points of submarine groundwater discharge into surface waters. With a 56 s half-life any detection of thoron must mean that one is close to a source. To improve measurement sensitivity, the influence of airflow rates on both radon and thoron measurements have been investigated using a RAD7 detector. Based on lab experiments, 2.5 L min−1represents an optimum airflow rate for measuring both radon and thoron. Field data from a coastal zone demonstrates that use of an external air pump produced a thoron response 65% higher than a standard setup.