[Imaging characteristics of PROPELLER T2-weighted imaging].
[Imaging characteristics of PROPELLER T2-weighted imaging].
复制标题
PROPELLER T2加权成像的成像特点[J].
DOI:
10.6009/jjrt.kj00003326583
复制
发表时间:
2004
期刊:
影响因子:
--
通讯作者:
K. Ohtomo
中科院分区:
文献类型:
--
作者:
M. Goto;S. Aoki;N. Hayashi;H. Mori;Yasushi Watanabe;Kenji Ino;Y. Satake;K. Nishida;Haruo Sato;Kyouhito Iida;K. Mima;K. Ohtomo
As the PROPELLER sequence is a combination of the radial scan and fast-spin-echo (FSE) sequence, it can be considered an FSE sequence with a motion correlation. However, there are some differences between PROPELLER and FSE owing to differences in k-space trajectory. We clarified the imaging characteristics of PROPELLER T2-weighted imaging (T2WI) for different parameters in comparison with usual FSE T2WI. When the same parameters were used, PROPELLER T2WI showed a higher signal-to-noise ratio (SNR) and lower spatial resolution than usual FSE. Effective echo time (TE) changed with different echo train lengths (ETL) or different bandwidths on PROPELLER, and imaging contrast changed accordingly to be more effective.