Evaluation of top, angle, and side cleaned FIB samples for TEM analysis
Evaluation of top, angle, and side cleaned FIB samples for TEM analysis
复制标题
DOI:
10.1002/jemt.20514
复制
发表时间:
2007-12-01
影响因子:
2.5
通讯作者:
van Tendeloo, Gustaaf
中科院分区:
文献类型:
--
作者:
Montoya, Eduardo;Bals, Sara;van Tendeloo, Gustaaf
ITEM specimens of a LaAlO3/SrTiO3 multilayer are prepared by FIB with internal lift out. Using a Ga+1 beam of 5 kV, a final cleaning step yielding top, top-angle, side, and bottom-angle cleaning is performed. Different cleaning procedures, which can be easily implemented in a dual beam FIB system, are described and compared; all cleaning types produce thin lamellae, useful for HRTEM and HAADF-STEM work up to atomic resolution. However, the top cleaned lamellae are strongly affected by the curtain effect. Top-angle cleaned specimens show an amorphous layer of around 5 nm at the specimen surfaces, due to damage and redeposition. Furthermore, it is observed that the LaAlO3 layers are preferentially destroyed and transformed into amorphous material, during the thinning process.