Screening with automated perimetry using a threshold-related three-level algorithm.
Screening with automated perimetry using a threshold-related three-level algorithm.
复制标题
使用阈值相关的三级算法通过自动视野检查进行筛查。
DOI:
10.1016/s0161-6420(86)33646-7
复制
发表时间:
1986
期刊:
影响因子:
13.7
通讯作者:
Auer,C
中科院分区:
文献类型:
--
作者:
Kosoko,O;Sommer,A;Auer,C
Eighty eyes were tested using the computer-driven, automated threshold-related, three-level screening strategy of the Humphrey Field Analyzer. Using Goldmann perimetry as the standard, different statistical algorithms were used in an attempt to differentiate eyes with glaucoma (n = 24) from normal controls (n = 56). Seventeen or more absolute and/or relative defects anywhere in the full field test area identified approximately 95% of the glaucoma eyes and screened out 80% of the control eyes. Retesting these criteria on a second group of 25 glaucomatous and 108 control eyes provided comparable results. The number of stimulus presentations can be substantially reduced, without deterioration in either sensitivity or specificity, by using a two-level strategy and limiting testing to the nasal field.