Multisite fMRI reproducibility of a motor task using identical MR systems
Multisite fMRI reproducibility of a motor task using identical MR systems
复制标题
DOI:
10.1002/jmri.21118
复制
发表时间:
2007-10-01
影响因子:
4.4
通讯作者:
Amaro, Edson, Jr.
中科院分区:
文献类型:
--
作者:
Costafreda, Sergi G.;Brammer, Michael J.;Amaro, Edson, Jr.
Purpose: To estimate between-scanner functional MRI (fMRI) reproducibility in a multisite study.Materials and Methods: A total of five identical 1.5T MR systems were used to repeatedly scan five subjects while performing a finger tapping task. A two-way (scanners, subjects) random effects analysis of variance (ANOVA) was used to estimate between-scanner and between-subject variability on two outcome variables: task-related mean blood oxygenation level dependent (BOLD) signal change and volume of activation within a motor region-of-interest (ROI).Results: Between-scanner variability of fMRI data accounted for a small proportion of the total variation in the BOLD signal change (8.34%, P = 0.114) and volume of activation (5.46%, P = 0.203). Between-subject variation accounted for more than half of the total variation for both measurements (57.17% and 54.46%, respectively, P < 0.01).Conclusion: These results support the feasibility of multisite studies using identical scanner systems.