A METHOD FOR CORRECTLY SETTING THE RF FLIP ANGLE
A METHOD FOR CORRECTLY SETTING THE RF FLIP ANGLE
复制标题
DOI:
10.1002/mrm.1910090104
复制
发表时间:
1989-01-01
影响因子:
3.3
通讯作者:
SANDSTROM, JC
中科院分区:
文献类型:
--
作者:
PERMAN, WH;BERNSTEIN, MA;SANDSTROM, JC
Currently the accepted method for setting the correct rf power levels to achieve 90.degree. and 180.degree. rf pulses for MR imaging is to peak the echo amplitude of a rf spin-echo sequence. The echo amplitude of this .alpha.-2.alpha. pulse sequence is proportional to sin3(.alpha.) and has a relatively broad maximum. Recently another method for setting the rf flip angle by maximizing the ratio of the stimulated echo to the primary echo amplitudes (in a 3.alpha. sequence) demonstrated accuracy similar to that of the spin-echo method using a shorter repetition time. We present a new, more sensitive, and more accurate method for setting the correct rf power levels for 90.degree. and 180.degree. rf pulses. In this method, based upon the stimulated echo pulse sequence, we are able to accurately set the rf power to within .+-. 0.1 dB by minimizing the signal amplitude of the third spin echo. This null method works for both selective and nonselective rf pulses of flip angle 90.degree. or 180.degree., allowing the user to accurately adjust the relative amplitudes of the four rf pulse types within a single pulse sequence.