INFLUENCE OF PROSTHESIS ALIGNMENT ON THE STANDING BALANCE OF BELOW-KNEE AMPUTEES

INFLUENCE OF PROSTHESIS ALIGNMENT ON THE STANDING BALANCE OF BELOW-KNEE AMPUTEES
复制标题

DOI:
10.1016/0268-0033(94)90008-6
复制
发表时间:
1994-07-01
影响因子:
1.8
通讯作者:
HAKIM, N
HAKIM, N
中科院分区:
工程技术3区
文献类型:
--
作者:
ISAKOV, E;MIZRAHI, J;HAKIM, N

文献摘要

被引文献

相似文献

The effect of alignment of the prosthesis on the standing balance of below knee amputees was investigated in this work. Alignment variations consisted of varus and valgus tilt and plantar- and dorsiflexions. The foot-ground reaction forces measured in the anteroposterior, mediolateral, and vertical directions as well as global parameters describing the total sway activity, asymmetry, and weight-bearing imbalance were used to compare the various alignment positions studied. The results indicate the presence of a common standing pattern in below-knee amputees, where activity of the foot-ground reaction forces in the anteroposterior direction is significantly higher in the contralateral limb compared with the amputated limb. It was also found that the varus and valgus alignments were the least stable of all alignments.RelevanceThis study has shown that below-knee amputees were able to adapt themselves to alignment changes and to keep their balance almost unaffected, even when standing took place under malaligned conditions. It demonstrates the uses and limits of postural sway parameters in the procedure of optimizing the alignment of below knee prostheses.