Characterizing contact impedance, signal quality and robustness as a function of the cardinality and arrangement of fingers on dry contact EEG electrodes
Characterizing contact impedance, signal quality and robustness as a function of the cardinality and arrangement of fingers on dry contact EEG electrodes
复制标题
将接触阻抗、信号质量和鲁棒性表征为干接触 EEG 电极上手指的基数和排列的函数
DOI:
10.1109/embc.2014.6944440
复制
发表时间:
2014
期刊:
影响因子:
--
通讯作者:
R. Jafari
中科院分区:
文献类型:
--
作者:
Viswam Nathan;R. Jafari
Continuous monitoring of patients' electroencephalography (EEG) outside of clinical settings will be valuable for detecting the onset of medical conditions such as epilepsy, as well as for enabling patients with physically disabling conditions like amyotrophic lateral sclerosis (ALS) to communicate using a brain-computer interface (BCI). This requires the development of a wearable dry-contact EEG system that takes into account not only the signal quality but also the robustness of the system for everyday use. To this end, we investigate whether certain designs of dry electrodes lend themselves to better characteristics overall with respect to these factors. Five different metallic finger-based dry electrodes were designed and scalp electrode impedance was used to compare them under varying capping conditions, followed by an evaluation of how well they captured steady state visually evoked potentials (SSVEP). Our findings indicate that configurations with a relatively low density of fingers can more effectively penetrate through hair on the scalp and are more robust to varying conditions. This was confirmed to be a statistically significant observation through a one-sided paired t-test that resulted in a p-value <; 0.004.