SCALING LIMITATIONS OF SILICON MULTICHANNEL RECORDING PROBES
SCALING LIMITATIONS OF SILICON MULTICHANNEL RECORDING PROBES
复制标题
DOI:
10.1109/10.43605
复制
发表时间:
1990-01-01
影响因子:
4.6
通讯作者:
WISE, KD
中科院分区:
文献类型:
--
作者:
NAJAFI, K;JI, J;WISE, KD
The scaling limitations of multichannel recording probes fabricated for use in neurophysiology using silicon integrated circuit technologies are described. Scaled silicon probe substrates 8- mu m thick and 16- mu m wide can be fabricated using boron etch-stop techniques. Theoretical expressions for calculating the thickness and width of silicon substrates have been derived and agree closely with experimental results. The effects of scaling probe dimensions on its strength and stiffness are described. The probe shank dimensions can be designed to vary the strength and stiffness for different applications. The scaled silicon substrates have a fracture stress of about 2*10/sup 10/ dyn/cm/sup 2/, which is about six times that of bulk silicon, and are strong and very flexible. Scaling the feature sizes of recording electrode arrays down to 1 mu m is possible with less than 1% electrical crosstalk between channels.<>