A calibration method for broad-bandwidth cavity enhanced absorption spectroscopy performed with supercontinuum radiation
A calibration method for broad-bandwidth cavity enhanced absorption spectroscopy performed with supercontinuum radiation
复制标题
DOI:
10.1007/s00340-010-4044-4
复制
发表时间:
2011-02
期刊:
影响因子:
--
通讯作者:
T. Laurila;I. Burns;J. Hult;J. H. Miller;C. Kaminski
中科院分区:
文献类型:
--
作者:
T. Laurila;I. Burns;J. Hult;J. H. Miller;C. Kaminski
An efficient calibration method has been developed for broad-bandwidth cavity enhanced absorption spectroscopy. The calibration is performed using phase shift cavity ring-down spectroscopy, which is conveniently implemented through use of an acousto-optic tunable filter (AOTF). The AOTF permits a narrowband portion of the SC spectrum to be scanned over the full high-reflectivity bandwidth of the cavity mirrors. After calibration the AOTF is switched off and broad-bandwidth CEAS can be performed with the same light source without any loss of alignment to the set-up. We demonstrate the merits of the method by probing transitions of oxygen molecules O2and collisional pairs of oxygen molecules (O2)2in the visible spectral range.