Sampling errors in projection reconstruction MRI.
Sampling errors in projection reconstruction MRI.
复制标题
投影重建 MRI 中的采样误差。
DOI:
10.1002/mrm.1910400317
复制
发表时间:
1998
影响因子:
3.3
通讯作者:
Joseph,PM
中科院分区:
文献类型:
--
作者:
Joseph,PM
Certain kinds of artifacts have been reported when projection reconstruction (PR) techniques are used in magnetic resonance imaging (MRI). These will occur if the spacing between samples ink‐space is too large. It has been suggested that PR requires finerk‐space sampling than does two‐dimensional Fourier reconstruction, and that the usual Nyquist criterion is inadequate. This paper examines this problem, with the conclusion that the Nyquist sampling criterion is adequate to avoid aliasing effects, provided that a band‐limited interpolation is used ink‐space. This procedure is motivated by analysis of the PR technique as it is commonly implemented in x‐ray computed tomography. A related problem is shown to be the construction of a filter function ink‐space that gives proper weight to the low spatial frequencies. It is shown that a simplekfilter does not satisfy this requirement, and a procedure for deriving a suitable filter is described. The methods are tested in simulated PR profiles of circular disks of two different sizes. It is shown that the combination of the two new methods gives virtually perfect reconstruction for disks up to the size implied by the Nyquist limit.