A phantom for diffusion-weighted MRI (DW-MRI)
A phantom for diffusion-weighted MRI (DW-MRI)
复制标题
DOI:
10.1002/jmri.23950
复制
发表时间:
2013-07-01
影响因子:
4.4
通讯作者:
Aboagye, Eric O.
中科院分区:
文献类型:
--
作者:
Lavdas, Ioannis;Behan, Kevin C.;Aboagye, Eric O.
Purpose: To develop tissue-equivalent diffusivity materials and build a spherical diffusion phantom which mimics the conditions typically found in biological tissues. Also, to assess the reproducibility of ADC measurements from a whole-body diffusion protocol.Materials and Methods: Nickel-doped agarose/sucrose gels were manufactured and used to build a spherical diffusion phantom with tissue-equivalent relaxation and diffusion compartments. The temporal stability of the gels was monitored for a period of 8 weeks and, using the same measurements, the reproducibility of ADC was assessed in a 1.5 Tesla (T) clinical system.Results: The temporal stability of the nickel-doped agarose/sucrose gels diffusion properties was excellent (average coefficient of variation [CV] for ADC in all phantom compartments = 1.27%). The average CV for ADC measurements, excluding the phantom compartments affected by artifacts, was 0.76% showing that the reproducibility of ADC measurements using an EPI DW-MRI protocol is very good.Conclusion: Nickel-doped agarose/sucrose gels can be used as reference materials for MRI diffusion measurements and show excellent short-term stability with respect to ADC. A phantom made of these materials can be invaluable in optimizing DW-MRI protocols, developing novel pulse sequences for DW-MRI, or comparing ADC values among field strengths, vendors, and imaging centers.