A novel device for measuring arterial stiffness using finger-toe pulse wave velocity: Validation study of the pOpmetre®
A novel device for measuring arterial stiffness using finger-toe pulse wave velocity: Validation study of the pOpmetre®
复制标题
DOI:
10.1016/j.acvd.2014.12.003
复制
发表时间:
2015-04-01
影响因子:
3
通讯作者:
Laurent, Stephane
中科院分区:
文献类型:
--
作者:
Alivon, Maureen;Thao Vo-Duc Phuong;Laurent, Stephane
Background. - The finger-toe pathway could be a good alternative for assessing arterial stiffness conveniently.Aim. - To evaluate the accuracy of the pOpmetre (R) - a new device that measures finger-toe pulse wave velocity (ft-PWV).Methods. - The pOpmetre has two photodiode sensors, positioned on the finger and the toe. Pulse waves are recorded continuously for 20 seconds, and the difference in pulse wave transit time between toe and finger (ft-TT) is calculated. The travelled distance is estimated using subject height. Study 1 compared ft-PWV with carotid-femoral PWV (cf-PWV) obtained by the reference method (SphygmoCor (R)) in 86 subjects (mean age 53 +/- 20 years), including 69 patients with various pathologies and 17 healthy normotensives. Study 2 compared changes in ft-PWV and cf-PWV during a cold pressor test in 10 healthy subjects. Study 3 assessed repeatability in 45 patients.Results. - ft-PWV correlated significantly with cf-PWV (R-2 = 0.43; P < 0.0001). A better correlation was found in terms of transit time (R-2 = 0.61; P < 0.0001). The discrepancy between transit times was related to age. The cold pressor test induced parallel changes in cf-PWV and ft-PWV, with increased aortic stiffness that was reversible during recovery. Intra-session repeatability was very good, with a coefficient of variation of 4.52%.Conclusion. - The pOpmetre (R) allows measurement of arterial stiffness in routine clinical practice. The greatest advantages of ft-PWV are simplicity, rapidity, feasibility, acceptability by patients and correct agreement with the reference technique. Further studies are needed to adjust for bias and to validate the pOpmetre in larger populations. (C) 2015 Elsevier Masson SAS. All rights reserved.