SIMULATED PULMONARY NODULES - DETECTION WITH DUAL-ENERGY DIGITAL VERSUS CONVENTIONAL RADIOGRAPHY
SIMULATED PULMONARY NODULES - DETECTION WITH DUAL-ENERGY DIGITAL VERSUS CONVENTIONAL RADIOGRAPHY
复制标题
DOI:
10.1148/radiology.160.3.3526398
复制
发表时间:
1986-09-01
期刊:
影响因子:
19.7
通讯作者:
BARNES, GT
中科院分区:
文献类型:
--
作者:
NIKLASON, LT;HICKEY, NM;BARNES, GT
Performance of a prototype dual-energy digital chest radiography unit in detaching calcified and noncalcified simulated pulmonary nodules was compared with that of a highly optimized, conventional system. Nodules ranging in size (0.5, 1.0, and 1.6 cm), in number (five to 11), and in calcium content (0-25 mg) were superimposed over the lungs of a frozen, unembalmed, human chest phantom. For each technique, six observers examined 50 posteroanterior projections with different randomized nodule locations. Detection consisted of locating and assigning a level of confidence to each perceived nodular opacity. The resulting plots of the true-positive fraction versus the mean number of false-positive calls per projection indicate that for both calcified and noncalcified nodules, the digital unit performed significantly better (P < .01).