Characterization measurement of a thick CdTe detector for BNCT-SPECT - Detection efficiency and energy resolution
Characterization measurement of a thick CdTe detector for BNCT-SPECT - Detection efficiency and energy resolution
复制标题
DOI:
10.1016/j.apradiso.2014.01.023
复制
发表时间:
2014-06-01
影响因子:
1.6
通讯作者:
Kato, Itsuro
中科院分区:
文献类型:
--
作者:
Murata, Isao;Nakamura, Soichiro;Kato, Itsuro
Author's group is carrying out development of BNCT-SPECT with CdTe device, which monitors the therapy effect of BNCT in real-time. From the design calculations, the dimensions were fixed to 1.5 x 2 x 30 mm(3). For the collimator it was confirmed that it would have a good spatial resolution and simultaneously the number of counts would be acceptably large. After producing the CdTe crystal, the characterization measurement was carried out. For the detection efficiency an excellent agreement between calculation and measurement was obtained. Also, the detector has a very good energy resolution so that gamma-rays of 478 key and 511 keV could be distinguished in the spectrum. (C) 2014 Elsevier Ltd. All rights reserved.