Monitoring Ravulizumab effect on complement assays
Monitoring Ravulizumab effect on complement assays
复制标题
DOI:
10.1016/j.jim.2020.112944
复制
发表时间:
2021-02-02
影响因子:
2.2
通讯作者:
Murray, David L.
中科院分区:
文献类型:
--
作者:
Willrich, Maria A., V;Ladwig, Paula M.;Murray, David L.
Ravulizumab is a new C5 inhibitor therapeutic monoclonal antibody with a longer half-life than eculizumab. Monitoring complete complement blockade by eculizumab has allowed personalized therapy in specific settings. Similar action is expected with ravulizumab. Ravulizumab has 4 different amino acids from eculizumab, which allow greater affinity for the FcRn immunoglobulin receptor and change the affinity of the molecule for C5. Here we investigate if clinical lab tests traditionally used to monitor complement blockade for eculizumab are appropriate for monitoring complement blockade caused by ravulizumab.De-identified serum samples with known normal complement activity were spiked with increasing amounts of ravulizumab, from zero to 1000 mu g/mL. Measurement of classical pathway function (CH50) and C5 function using a liposome method (Wako Diagnostics) showed >50% complement inhibition starting with 50 mu g/mL of ravulizumab, but inhibition >95% of complement activity was not achieved, with residual measurements of 11% at 700 mu g/mL. In contrast, measurement of alternative pathway function using an ELISA (AH50, Wieslab) showed alternative pathway function inhibition of 80% at 50 mu g/mL of ravulizumab and > 95% at 200 mu g/mL, which is consistent with expected therapeutic concentrations of ravulizumab >175 mu g/mL. If replicated in patient sera, AH50 could be a suitable therapeutic monitoring tool.