In vitro comparison of the charge-injection limits of activated iridium oxide (AIROF) and platinum-iridium microelectrodes
In vitro comparison of the charge-injection limits of activated iridium oxide (AIROF) and platinum-iridium microelectrodes
复制标题
DOI:
10.1109/tbme.2005.851503
复制
发表时间:
2005-09-01
影响因子:
4.6
通讯作者:
Plante, TD
中科院分区:
文献类型:
--
作者:
Cogan, SF;Troyk, PR;Plante, TD
The charge-injection limits of activated iridium oxide electrodes (AIROF) and PtIr microelectrodes with similar geometric area and shape have been compared in vitro using a stimulation waveform that delivers cathodal current pulses with current-limited control of the electrode bias potential in the interpulse period. Charge-injection limits were compared over a bias range of 0.1-0.7 V (versus Ag vertical bar AgCl) and pulse frequencies of 20, 50, and 100 Hz. The AIROF was capable of injecting between 4 and 10 times the charge of the PtIr electrode, with a maximum value of 3.9 mC/cm(2) obtained at a 0.7 V bias and 20 Hz frequency.