Body composition analysis by dual energy X-ray absorptiometry in female diabetics differ between manufacturers

Body composition analysis by dual energy X-ray absorptiometry in female diabetics differ between manufacturers
复制标题

DOI:
10.1038/sj.ejcn.1600424
复制
发表时间:
1997-07-01
影响因子:
4.7
通讯作者:
Svendsen, OL
Svendsen, OL
中科院分区:
医学3区
文献类型:
--
作者:
Kistorp, CN;Svendsen, OL

文献摘要

被引文献

相似文献

目的:比较两种双能X射线吸收测定法(DXA)仪器(即Hologic Inc的QDR-2000和Lunar的DPX)在瘦型和肥胖型受试者亚群中的身体组成结果。设计:对85名女性糖尿病患者进行的横断面研究(BMI 18-43 kg/m(2))。结果:脂肪组织质量(FTM)、脂肪百分比(FAT%)和全身骨矿物质含量(TBMC)的回归线与同一直线有显著性差异(P < 0.01),而瘦肉组织质量(LTM)无显著性差异(P> 0.05)。然而,相关性很高(r(2)> 0.95),相应的SEE%很低(0.8-4.8%),并且与BMI无关。QDR-2000测量的FTM和FAT%比DPX高10%,LTM和TBMC低6%(P < 0.001)。两个系统的单个结果无法直接比较。DXA测量人体成分的标准化是非常必要的。
Objective: Comparison of body composition results by two dual energy X-ray absorptiometry (DXA) instruments, namely QDR-2000 from Hologic Inc and DPX from Lunar in subpopulations of lean and obese subjects.Design: Cross-sectional study with 85 female diabetics (BMI 18-43 kg/m(2)) measured with both DXA instruments.Results: The regression lines for fat tissue mass (FTM), FAT% and total body bone mineral content (TBMC), but not lean tissue mass (LTM), were different from the line of identity (P < 0.01). However, the relationships were high (r(2) > 0.95), and the corresponding SEE%'s were low (0.8-4.8%), and were independent of BMI. FTM and FAT% measured by the QDR-2000 were 10% higher, and LTM and TBMC 6% lower, than by DPX (P < 0.001).Conclusions: There were lack of agreements between total body composition results by DPX Lunar, and QDR-2000 Hologic Inc. Individual results on the two systems cannot be directly compared. Standardization of body composition measurements by DXA is strongly needed.